Article No. VK-71000, VK-71010

Ultrasound Pulser Receiver SF710 resp. SF710sc

Transmitter and receiver for ultrasound pulse waves

High-Performance Ultrasound Technology

The Ultrasound Pulser Receiver SF710 and SF710sc from GAMPT are advanced instruments designed for high-frequency ultrasound applications. With powerful spike pulses up to 400 V, low-noise amplification, and intuitive software integration, they deliver reliable performance for research, development, and industrial testing.

Key features

  • Powerful Signal Generation – Up to 400 V spike pulses for precise excitation of ultrasound transducers
  • Flexible Measurement Modes – Reflection, transmission, and hydrophone calibration (SF710sc)
  • Wide Frequency Range – Receiver bandwidth from 40 kHz to 20 MHz for diverse applications
  • Advanced Software Tools – Seamless control and signal analysis with GS-EchoView (FFT, Cepstrum, A-scan)
  • Hydrophone Calibration – Exclusive to the SF710sc, enabling secondary calibration up to 100 MHz.

Applications

  • Non-Destructive Testing (NDT): Detect material flaws, analyze thin layers, and ensure product safety
  • Ultrasound Transducer Development: Characterize and optimize performance with precision
  • Material Science: Measure sound velocity, acoustic impedance, and absorption properties
  • Medical Ultrasound Research: Support device validation, calibration, and quality assurance
  • Hydrophone Testing & Calibration: Perform accurate secondary calibration with the SF710sc

Pulser

  SF710 SF710sc
Article-No. VK-71000 VK-71010
Pulse type Pulser Sensor 1 T/R Pulse Calib T/R
Pulse signal spike (Dirac), negative spike (Dirac), positive
Rise time 10 ns 15 ns
Pulse voltage max. 400 V 400 V
Voltage step 0, 10, 30, 100, 200, 300, 400 V 0, 10, 30, 100, 200, 300, 400 V
Pulse width < 30 ns < 40 ns


Receiver
Amplifier: 2 stage broadband, max. 65 dB
Bandwidth: 40 kHz – 20 MHz, -3 dB
Gain: 0, 5, 10, 15, 20, 25, 30, 35 dB
TGC (duration-dependent amplification): 0 – 30 dB

Modes
Pulser: normal, calibration
Measurement: reflection, transmission
PRF: 10 Hz – 2 kHz
TGC: threshold 0 – 30 dB, wide 5 – 130 µs, slope 0 – 30 dB, start 5 – 135 µs
Trigger: internal, external

Connections on the front panel
Sensor 1: signal output/input, max. 400 V out
Sensor 2: signal input
Calib: signal output/input, max. 400 V out
Trigger, TGC, RF-Signal, A-Scan: signal output, BNC

Connections on the back panel
Trigger: 1 – 5 V input, BNC
USB Type B

 

 

  • Ultrasound pulser receiver SF710 resp. SF710sc
  • USB cable
  • Ground cable
  • Conductivity temperature controller (optional)
  • Software (SFS91xx, GS-EchoView*.dll)