Article No. (see product description)

Standard membrane hydrophone

MH-SD-0.2 | MH-SB-0.2 | MH-SD-0.4 | MH-SD-0.5 | MH-SD-1.0 | MH-CC-0.4

Standard membrane hydrophone MH-SD-0.2

Gold standard hydrophone – precision for the highest demands
The MH-SD-0.2 sets standards in quality and reliability. It complies with all relevant IEC standards, thus meeting international standards and offering maximum safety in use. Thanks to its high long-term stability, it guarantees precise measurement results even under the most demanding conditions.

VK-11025
Technical data
  • active diameter: 200 µm
  • sensitivity (typical): 1.0*10-7 V/Pa @5 MHz
  • max. pressure: 10 MPa
  • usable frequency range: 0.5-100 MHz
  • frequency response: +/-2 dB @1-40 MHz
  • membrane diameter: 80 mm
  • outer diameter: 130 mm
  • height: 15 mm
  • min. measure distance: 6 mm front, 9 mm back
Fields of application
  • standard sound field measurements
  • regulatory measurements according to IEC 62127 and 60601 for all medical Ultrasound devices
  • characterisation of NDT sensors
Users
  • certification laboratories
  • manufacturers of medical ultrasound devices
  • research and development laboratories
  • universities

Standard membrane hydrophone with backing MH-SB-0.2

Highest mechanical stability – recommended for fast scanning processes
The MH-SB-0.2 impresses with its exceptional mechanical robustness. This makes it ideal for fast scanning processes where precision and reliability are crucial.

VK-11026  
Technical data
  • active diameter: 200 µm
  • sensitivity (typical): 1.0*10-7 V/Pa @5 MHz
  • max. pressure: 10 MPa
  • usable frequency range: 0.5-100 MHz
  • frequency response: +/-2 dB @1-40 MHz
  • membrane diameter: 75 mm
  • outer diameter: 130 mm
  • height: 15 mm
  • min. measure distance: 2.5 mm, front only
Fields of application
  • continuous sound field measurements
  • field measurements with high speed positioning
  • quality measurements of ultrasound sensors
Users
  • certification laboratories
  • manufacturers of medical ultrasound devices

Standard membrane hydrophone MH-SD-0.4

Proven standards – strong performance
The MH-SD-0.4 is manufactured according to older hydrophone standards and impresses with its high signal strength and low noise. This gives you reliable measurement results with clear data quality.

VK-11041  
Technical data
  • active diameter: 400 µm
  • sensitivity (typical): 1.0*10-7 V/Pa @5 MHz
  • max. pressure: 10 MPa
  • usable frequency range: 0.5-40 MHz
  • frequency response: +/-2 dB @1-20 MHz
  • membrane diameter: 80 mm
  • outer diameter: 130 mm
  • height: 15 mm
  • min. measure distance: 6 mm front, 9 mm back
Fields of application
  • broadband measurements with low pressure amplitudes
  • regulatory measurements according to older standards for diagnostic ultrasound devices
User
  • certification laboratories
  • manufacturers of medical ultrasound devices

Standard membrane hydrophone MH-SD-0.5

Proven standards – strong performance
The MH-SD-0.5 is manufactured according to older hydrophone standards and impresses with its high signal strength and low noise. This gives you reliable measurement results with clear data quality.

VK-11043  
Technical data
  • active diameter: 500 µm
  • sensitivity (typical): 1.0*10-7 V/Pa @5 MHz
  • max. pressure: 10 MPa
  • usable frequency range: 0.5-40 MHz
  • frequency response: +/-2 dB @1-20 MHz
  • membrane diameter: 80 mm
  • outer diameter: 130 mm
  • height: 15 mm
  • min. measure distance: 6 mm front, 9 mm back
Fields of application
  • broadband measurements with low pressure amplitudes
  • regulatory measurements according to older standards for diagnostic ultrasound devices
User
  • certification laboratories
  • manufacturers of medical ultrasound devices

Standard membrane hydrophone MH-SD-1.0

Proven standards – strong performance
The MH-SD-1.0 is manufactured according to older hydrophone standards and impresses with its high signal strength and low noise. This gives you reliable measurement results with clear data quality.

VK-11042  
Technical data
  • active diameter: 1000 µm
  • sensitivity (typical): 2.4*10-7 V/Pa @5 MHz
  • max. pressure: 10 MPa
  • usable frequency range: 0.5-30 MHz
  • frequency response: +/-4 dB @1-20 MHz
  • membrane diameter: 80 mm
  • outer diameter: 130 mm
  • height: 15 mm
  • min. measure distance: 6 mm front, 9 mm back
Fields of application
  • broadband measurements with very low pressure amplitudes
  • regulatory measurements according to older standards for diagnostic Ultrasound devices
  • measurements for acoustic material characterization (absorption measurement)
Users
  • certification laboratories
  • manufacturers of medical ultrasound devices
  • research labs
  • universities

Chemical capsulated membrane hydrophon MH-CC-0.4

Robust and resistant
With covered electrodes and a chemical-resistant housing material, the MH-CC-0.4 is specially designed for use in aggressive media such as acids. This ensures durability and reliable performance even under extreme conditions.

VK-11031  
Technical data
  • active diameter: 325 µm
  • sensitivity (typical): 8.0*10-8 V/Pa @5 MHz
  • max. pressure: 20 MPa
  • usable frequency range: 0.5-30 MHz
  • frequency response: +/-4 dB @1-20 MHz
  • membrane diameter: 50 mm
  • outer diameter: 130 mm
  • height: 18 mm
  • min. measure distance: 8 mm front, 10 mm back
Fields fo application
  • measurements in aggressive liquids
  • development of sensors for chemical, petrol or food industries
  • development of flow sensors
  • quality management during sensor manufacturing
Users
  • manufacturers of ultrasound sensors
  • research labs


(Data sheets available on request)